SmartReader is designed to keep users informed using a series of alerts and Suspect warnings.

SmartReader Alert

Scanned Too Soon

 

 

This alert may be displayed the user does not allow enough time for the rapid drug test kit to develop.

 

SOLUTION:  Wait at least 15 seconds and rescan.

System Alerts

SmartReader Alert

Poor Lighting

 

 

This alert may be displayed when a shadow, too little or too much light is preventing an accurate test reading.

 

SOLUTION:  Rotate the stand slightly left or right

System Alerts

System Alerts

SmartReader Alert

Test Kit unfamiliar

 

This alert may be displayed when a QR code is scanned without a test kit or placed on the wrong test kit.

SOLUTION:  Use approved test kits only. This warning triggers an investigation.

SmartReader Alert

Bad Test Position

 

 

This alert may be displayed when the test is placed on the stand incorrectly or a smart stand is missing.

 

SOLUTION:  Place the test kit on the stand properly

System Alerts

Sample Double-Check Alert

Adulterated

 

 

This warning may appear when the donor sample color appears abnormal.

 

SOLUTION:  Run an SVT (specimen validity test) or discard sample and recollect.

Sample Double-Check Alert

Overheated

 

 

This warning may appear when the test lines are fully developed at 98˚ or higher.

 

SOLUTION:  Discard sample and request a new donor sample.

Sample Double-Check Alert

Underdeveloped

 

 

This warning will appear when the line development is abnormally light across multiple test strips.

 

SOLUTION:  Wait for 3-minutes and rescan 

Suspect Alerts

Suspect Alerts

Suspect Alerts

901 Port Centre Parkway, STE 8

Portsmouth, VA 23704

877 868-9967